Abstract
The SASQART practice guidelines were set forth by the South African Medical Physics Society & South African Association of Physicists in Medicine and Biology. The intended purpose of this document and the individual topic-specific sections is to provide guidelines as to a minimum set of tests, result tolerances and a minimum frequency of these tests to be performed by medical physicists and radiation therapy staff in a radiation oncology department to promote patient safety. The tests, tolerances and frequencies from the international references have been adapted to realistically reflect the South African working environment and resource availability, while maintaining the highest clinical and scientific quality of care achievable.
Contribution: Replaces the 2014 guidelines, establishing recommended quality assurance standards for radiation oncology in South Africa.
Keywords: SASQART; radiotherapy; quality assurance; Linac; dosimetry; radiation protection.
Introduction
In November 2021, a task group was appointed by the South African Medical Physics Society (SAMPS), a member society of the South African Association of Physicists in Medicine and Biology (SAAPMB). This task team was chosen to be representative of multiple provinces in the country, private, academic and public radiotherapy facilities, and to incorporate medical physicists with different areas of expertise. The task group was given a mandate to produce a document that was intended to update or replace the existing 2014 SASQART guidelines that are referred to in the South African Health Products Regulatory Authority (SAHPRA) license conditions for medical radiotherapy devices. The document was intended to cover all forms of quality assurance (QA) to be performed in a clinical Radiation Oncology Department at the time of the formation of the task group.
The intended purpose of this document and the individual topic-specific sections is to provide guidelines as to a minimum set of tests and tolerances, and a minimum frequency of these tests to be performed by medical physicists and radiation therapy staff in radiotherapy to promote patient safety. The tests and test frequencies were adapted based on both local experience and additional international guidelines, such as those from Canadian Association of Provincial Cancer Agencies (CAPCA), American Association of Physicists in Medicine (AAPM), International Atomic Energy Agency (IAEA), Institute of Physics and Engineering in Medicine (IPEM) and Netherlands Commission on Radiation Dosimetry (NCS). The CAPCA gave permission for us to use their Standards for Quality Control as a basis for the original 2014 SASQART documents. The assistance and information sharing of CAPCA is appreciated but they are not in any way responsible for the final SASQART documentation. The test frequencies from the international references have been adapted to realistically reflect the South African working environment, while ensuring patient safety. There was a long process of consultation between task team members, and the comments from the South African medical physics community were incorporated after the first round of edits in 2023. The final documents have been agreed upon by all task team members as recommendations to be proposed to SAMPS.
These tests and their frequency are unique to the environment of radiotherapy and should not be applied where the same equipment is used in other disciplines. For example, the tests for a computed tomography (CT) scanner are limited to the requirements for imaging for radiotherapy and do not cover the more rigorous demands of a CT scanner used in diagnostic radiology.
This document is only to be used as a guideline and not as a legally binding document for licensing conditions. If any clinic can show that their equipment shows historical stability beyond the frequency of testing suggested, then the professionalism of the medical physicist should allow for a reduction in the test frequency. In contrast, where equipment is known to have inherent instability below ‘normal’ standards, the medical physicist should perform tests at shorter intervals than suggested in this document.
The Radiation Protection document has been specifically designed for implementation by the Department of Health and SAHPRA to ensure radiation safety.
The techniques employed to establish the compliance of a particular test to the tolerances suggested in this document remain totally within the preference of the medical physicist, who should be able to explain the methodology adequately. It is beyond the scope of this document to prescribe experimental techniques and vendor-specific equipment. This should be left to the professional judgement of the physicist at each clinic. If tolerance cannot be reached because of economic, vendor, mechanical or electronic constraints, there should be documentation to show how the clinical procedures are adapted to accommodate these anomalies, without impacting upon patient safety. These tests should be conducted by a qualified medical physicist or radiotherapy personnel trained by a qualified medical physicist. Any test exceeding tolerance should be reported to a qualified medical physicist to determine the course of action.
The format of this document specifies tests which need to be performed within certain time periods. These time periods are depicted in Table 1.
TABLE 1: Definitions of frequency nomenclature. |
The tests performed during commissioning may be repeated during the routine quality control intervals. For quantities that can be measured, the ‘performance’ of a test is compared to an action and tolerance level. If the difference between the measured and expected value is at or below the tolerance level, no action is required. If the difference is larger than the ‘action’ level, immediate action is required. If this is not immediately possible, the use of the system must be restricted to clinical situations in which the identified inadequate performance is of no, or acceptable and understood, clinical significance.
Hein Fourie
Chair: SASQART Task Group
Christoph Trauernicht
Chairperson: SAMPS
Cobalt-60 teletherapy units
TABLE 2: Quality assurance tests for Cobalt-60 teletherapy units. |
Kilovoltage x-ray radiotherapy machines and Intra-Operative Radiotherapy
TABLE 3: Quality assurance tests for external kilovoltage x-ray therapy machines. |
TABLE 4: Quality assurance tests for internal kilovoltage x-ray therapy machines. |
CT scanners and CT-simulators
TABLE 5: Quality assurance tests for Computed Tomography (CT) scanners and CT-simulators. |
Dosimetry equipment
Devices for absolute dosimetry
TABLE 6: Quality assurance tests for secondary standard chambers and electrometers. |
TABLE 7: Quality assurance tests for field standard chambers and electrometer. |
Devices for relative dosimetry
TABLE 8: Quality assurance tests for in vivo Tthermoluminescent dosimeters (TLD) systems. |
TABLE 9: Quality assurance tests for film dosimetry. |
TABLE 10: Quality assurance tests for in vivo diode systems. |
TABLE 11: Quality assurance tests for automated beam scanning systems. |
Quality assurance devices
TABLE 12: Quality assurance tests for diode arrays. |
TABLE 13: Quality assurance tests for general phantom materials. |
TABLE 14: Quality assurance tests for thermometers and barometers. |
TABLE 15: Quality assurance tests for spirit levels and self-levelling lasers. |
TABLE 16: Quality assurance tests for distance measuring devices. |
TABLE 17: Quality assurance tests for radiation survey meters. |
Brachytherapy remote afterloaders
TABLE 18: Quality assurance tests for High Dose Rate (HDR), Pulse Dose Rate (PDR) and Low Dose Rate (LDR) remote afterloaders. |
TABLE 18 (Continues...): Quality assurance tests for High Dose Rate (HDR), Pulse Dose Rate (PDR) and Low Dose Rate (LDR) remote afterloaders. |
Megavoltage electronic portal imaging devices
TABLE 19: Quality assurance tests for Electronic Portal Imaging Device (EPID). |
TABLE 20: Quality assurance tests for Electronic Portal Imaging Device (EPID) dosimetry. |
Image guided radiotherapy
TABLE 21: Quality assurance tests for kV-based imaging. |
TABLE 22: Half Value Layer (HVL) values for x-ray tubes (Directorate: Radiation Control, Diagnostic QC). |
Intensity modulated radiotherapy and Volumetric modulated arc therapy
TABLE 23: Quality assurance tests for Intensity Modulated Radiotherapy and Volumetric Modulated Arc Therapy. |
TABLE 23 (Continues...): Quality assurance tests for Intensity Modulated Radiotherapy and Volumetric Modulated Arc Therapy. |
Medical linear accelerators
TABLE 24: Quality assurance tests for Linacs. |
Multileaf collimators
TABLE 25: Quality assurance tests for Multileaf Collimators (MLCs). |
Radiation protection
TABLE 26: Quality assurance tests for radiation protection in oncology. |
Therapeutic radioisotope administration
TABLE 27: Quality assurance tests for therapeutic radionuclides. |
Record and verify systems
TABLE 28: Quality assurance tests for Record and Verify (R&V) systems. |
Conventional radiotherapy simulators
TABLE 29: Quality assurance tests for radiotherapy simulators. |
Stereotactic radiosurgery/radiotherapy – Gamma knife based
TABLE 30: Quality assurance tests for Gamma Knife units. |
Stereotactic radiosurgery/radiotherapy – Linac based
TABLE 31: Quality assurance tests for linac-based Stereotactic Radiosurgery/Stereotactic Radiotherapy. |
Treatment planning systems
Treatment planning system auxiliary equipment
TABLE 33: Quality assurance tests for auxiliary Treatment Planning System (TPS) equipment. |
Bore based medical linear accelerators (such as halcyon, tomotherapy and MRI-linacs)
TABLE 34: Quality assurance tests for bore-based linacs. |
Acknowledgements
The SASQART task group would like to acknowledge and thank the following persons who authored the original 2014 SASQART guideline documents: Emma Viviers, Graeme Lazarus, Nanette Willemse Joubert, Freek du Plessis and Hannelie MacGregor.
Competing interests
The authors declare that they have no financial or personal relationships that may have inappropriately influenced them in writing this article. The author, C.J.T., serves as an editorial board member of this journal. The peer review process for this submission was handled independently, and the author had no involvement in the editorial decision-making process for this manuscript. The authors have no other competing interests to declare.
Authors’ contributions
All authors contributed equally to the writing, review and editing of the initial draft(s). H.F. and C.J.T. were lead contributors to the final review and editing, and project administration.
Ethical considerations
This article followed all ethical standards for research without direct contact with human or animal subjects.
Funding information
The article processing fee was generously covered by the South African Association for Physicist in Medicine and Biology.
Data availability
Data sharing is not applicable to this article as no new data were created or analysed in this study.
Disclaimer
The views and opinions expressed in this article are those of the authors and are the product of professional research. The article does not necessarily reflect the official policy or position of any affiliated institution, funder, agency or that of the publisher. The authors are responsible for this article’s results, findings and content.
References
- Institute of Physics and Engineering in Medicine. Physics aspects of quality control in radiotherapy. York: Institute of Physics and Engineering in Medicine; 1999.
- American Association of Physicists in Medicine. Report 46: Comprehensive QA for radiation oncology. Med Phys. 1994;21(4):581–618.
- Canadian Partnership for Quality Radiotherapy. Technical quality control guidelines for computed tomography simulators. Québec: CPQR; 2017.
- American Association of Physicists in Medicine. TG-66: Quality assurance for computed-tomography simulators and the computed tomography-simulation process. Med Phys. 2003;30(10):2762–2792.
- American Association of Physicists in Medicine. TG-233: Performance evaluation of computed tomography systems. Alexandria: American Association of Physicists in Medicine; 2019.
- American College of Radiology. Computed tomogrpahy quality control manual. Reston: American College of Radiology; 2017.
- International Atomic Energy Agency. HHS-19: Quality assurance programme for computed tomography: Diagnostic and therapy applications. Vienna: International Atomic Energy Agency; 2012.
- South African Health Products Regulatory Authority. Guidelines on requirements for licence holders with respect to quality control tests for diagnostic X-ray imaging systems. Johannesburg: South African Health Products Regulatory Authority; 2022.
- American Association of Physicists in Medicine. TG-142 report: Quality assurance of medical accelerators. Med Phys. 2009;36:4197–4212.
- American Association of Physicists in Medicine. TG-179: Quality assurance for image-guided radiation therapy utilizing CT-based technologies. Med Phys. 2012;39(4):1946–1963.
- American Association of Physicists in Medicine. AAPM medical physics practice guideline 2.a: Commissioning and quality assurance of X-ray–based image-guided radiotherapy systems. J Appl Clin Med Phys. 2014;15(1):3–13.
- Canadian Partnership for Quality Radiotherapy. Technical quality control guidelines for accelerator-integrated cone-beam systems for verification imaging. Toronto: Canadian Partnership for Quality Radiotherapy; 2015.
- American College of Radiology – American Society for Radiation Oncology. ACR–ASTRO practice parameter for image-guided radiation therapy (IGRT). Am J Clin Oncol. 2020;43(7):459–468. https://doi.org/10.1097/COC.0000000000000697
- Nederlandse Commissie Voor Stralingsdosimetrie. Report 32: Quality assurance of cone-beam CT for radiotherapy. Delft: NCS; 2019.
- Van Esch A, Huyskens DP, Behrens CF, et al. Implementing RapidArc into clinical routine: A comprehensive program from machine QA to TPS validation and patient QA. Med Phys. 2011;38(9):5146–5166.
- Bedford JL, Warrington AP. Commissioning of volumetric modulated Arc therapy (VMAT). Int J Radiat Oncol Biol Phys. 2009;73(2):537–545. https://doi.org/10.1016/j.ijrobp.2008.08.055
- American Association of Physicists in Medicine. TG 218: Tolerance limits and methodologies for IMRT measurement-based verification QA. Med Phys. 2018;45(4):e53–e83.
- Boonzaier WP. A multi-institutional quantitative survey of multi-leaf collimator accuracy using a digital picket fence test with sub-millimeter detection capabilities. Stellenbosch: Stellenbosch University; 2022.
- NCRP. Report 147: Structual shielding design for medical X-ray imaging facilities. Bethesda: NCRP; 2004.
- NCRP. Report 151: Structural shielding design and evaluation for megavoltage X- and gamma-ray radiotherapy facilities. 2005.
- South African Health Products Regulatory Authority. Guidelines for users of electronic therapeutic devices emitting ionizing radiation. Pretoria: SAHPRA; 2023.
- American Association of Physicists in Medicine. TG 275: Strategies for effective physics plan and chart review in radiation therapy. Med Phys. 2020;47(6):e236–e272.
- Mack A, Mack G, Scheib S, et al. Quality assurance in stereotactic radiosurgery/radiotherapy according to DIN 6875-1. Stereotactic Funct Neurosurg. 2005;82:235–243. https://doi.org/10.1159/000083175
- Canadian Association of Provincial Cancer Agencies. Stereotactic radiosurgery/radiotherapy – Standards for quality control at Canadian Radiation Treatment Centres. Toronto: CAPCA; 2006.
- Benedict SH, Yenice KM, Followill D, et al. Stereotactic body radiation therapy: The report of AAPM Task Group 101. Med Phys. 2010;37(8):4078–4101.
- Klein EE, Hanley J, Baymouth J, et al. Task Group 142 report: Quality assurance of medical accelerators. Med Phys. 2009;36(9):4197–4212.
- Netherlands Commission on Radiation Dosimetry. Process management and quality assurance for intracranial stereotactic treatment: Report 25. Delft: NCS; 2015.
|